An Equipment Monitor that is designed to monitor the performance of vacuum-assisted 132C-134C (270F - 273F) steam sterilizers each day.
As an Equipment Monitor, the 3M™ Comply™ Bowie-Dick Test Pack #1233 is designed to monitor the performance of vacuum-assisted 132C-134C (270F - 273F) steam sterilizers each day. Test packs will reproducibly detect residual air in the sterilizer chamber due to air leaks, air removal failures and inadequate vacuum - any of which can compromise sterility. Complies to ANSI/AAMI/ISO 11140-5:2007 Class 2.
Brands | Comply™ |
Model | 1233LF |
Product Type | Bowie-Dick |
Product Use | Equipment Monitoring |
Sterilisation Method | Steam |
Units per Case | 30.0 |
Overall Height (Metric) | 13 cm |
Overall Length (Metric) | 11 cm |
Overall Width (Metric) | 2 cm |
Safety Data Sheets
Document Number: MSDS_2712016_AU_EN
(PDF)